HiRes
series is a cooled digital cameras series born to detect extremely
faint light emissions.
There are two versions: the standard HiRes IV and the upgraded
model HiRes IV Plus, both of them made of two main components:
the head (with CCD sensor, Double stage Peltier cooling system,
electronics) and the controller, that includes the power supply.
This series displays KODAK and E2V Technologies sensors, either
marked by high definition (small pixels : <= 16 mm) or high
sensitivity (bigger pixels : >= 20 mm). They belong to the
front-illuminated kind, with thin indium electrodes able to raise
the Quantum Efficiency in the case of shorter wavelengths, or
to the back-illuminated type to make the most of the latest CCD
technology.
Among the
main application fields : Biology, Astronomy, Spectroscopy, RAMAN
Spectroscopy, Semiconductor Physics, Plasma Physics and X-Ray
Physics.
A line completely devoted to Spectroscopy is available.
HiRes IV also includes the former Hurricane specs.
For the HiRes
IV cameras we have appealed to sophisticated sampling and amplification
techniques as well as to a 16-bit A/D Converter, to be fully exploited
in their dynamics. Furthermore, the Double Stage Peltier cooling
system allows to perform DT= -50 °C.
All cameras have been equipped with integrated peripherals, such
as parallel ports and timers, allowing one to carry out a complete
automation of one's experiment.
Herewith below the specs of the cameras as well as the mostly
used sensors.
However: feel
free to ask for any other KODAK or E2V technologies sensor not
included in the following tables.
Specifications
(high-resolution)
| MODEL
|
CCD
|
ARRAY
FORMAT (HxV) |
PIXEL
(HxV) um |
SIZE
(HxV) mm |
FWC
ke- |
DARK
CURRENT
e-/pix.s@-5 °C |
Qe
@ 450,550,650 nm |
FILL
FACTOR % |
| HR400LE |
KAF-0401LE |
768x512
|
9x9 |
6.9x4.6
|
100 |
0.5 |
20, 35,
42 |
70 |
HR400E
HR400ME |
KAF-0402E
KAF-0402ME |
768x512
|
9x9 |
6.9x4.6
|
100 |
0.5 |
35, 55,
68
55, 67, 82 |
100 |
HR1300E
HR1300LE |
KAF-1301E
KAF-1301LE |
1280x1024
|
16x16 |
20.48x16.38
|
220
120 |
2.8
4.5 |
40, 55,
63
20, 32, 28 |
100
70 |
HR1600E
HR1600LE |
KAF-1602E
KAF-1602LE |
1536x1024
|
9x9 |
13.8x9.2
|
100
50 |
0.6
0.5 |
40, 50,
60
20, 35, 42 |
100
70 |
HR3200E
HR3200ME |
KAF-3200E
KAF3200ME |
2184x1472
|
6.8x6.8
|
14.85x10.26
|
55 |
0.5 |
40, 52,
65
60, 75, 82 |
100 |
HR6000E
HR6000LE |
KAF-6303E
KAF-6302LE |
3072x2048
3072x2034 |
9x9 |
27.65x18.48
|
100
50 |
0.5 |
40, 55,
64
9, 15, 13 |
100
70 |
High-Sensitivity
| MODEL
|
CCD
|
ARRAY
FORMAT (HxV) |
PIXEL
(HxV) um |
SIZE
(HxV) mm |
FWC
ke- |
DARK
CURRENT
e-/pix.s@-5 °C |
Qe
@ 450,550,650 nm |
FILL
FACTOR % |
| HR260E
|
KAF-0261E |
512x512 |
20x20 |
10.2x10.2 |
500 |
12.5 |
36, 52,
58 |
100 |
| HR1000E |
KAF-1001E |
1024x1024
|
24x24 |
24.5x24.5 |
500 |
17.2 |
40, 70,
63 |
100 |
| HR4300E
|
KAF-4301E |
2084x2084 |
24x24 |
50.0x50.0 |
570 |
4.5 |
40, 55,
65 |
100 |
HR4710F
HR4710B |
CCD47-10F
CCD47-10B |
1024x1024 |
13x13 |
13.3x13.3 |
100 |
4.4
11 |
8, 24,
40
82, 81, 76 |
100 |
HR7700F
HR7700B |
CCD7700F
CCD7700B |
512x512 |
24x24 |
12.3x12.3
|
300 |
8.5
29.5 |
8, 31,
45
82, 81, 76 |
100 |
Spectroscopy
| MODEL
|
CCD
|
ARRAY
FORMAT (HxV) |
PIXEL
(HxV) um |
SIZE
(HxV) mm |
FWC
ke- |
DARK
CURRENT
e-/pix.s@-5 °C |
Qe
@ 450,550,650 nm |
FILL
FACTOR % |
HR3011F
HR3011B |
CCD30-11F
CCD30-11B |
1024x256 |
26x26 |
26.6x6.7 |
500 |
11
43 |
21, 43,
45
82, 81, 76 |
100 |
HR4210F
HR4210B |
CCD42-10F
CCD42-10B |
2048x512 |
13.5x13.5 |
27.6x6.9 |
100 |
4.4
11 |
8, 32,
45
84, 82, 76 |
100 |
"E"
is the Enhanced (ITO) version, "F" means front-illuminated, "B"
back-illuminated.
The values concerning Quantum Efficiency, Full Well Capacity and
Dark Current have to be considered as typical.
| Readout
speed |
Up to 100
kpixel/s |
| Filter
wheel (opt.) |
External
|
| A/D Converter
|
16 bit
|
| Max total
noise |
5e-/15e-
|
| Partial
CCD Reading |
Programmable
|
| Binning
|
From 1x1
to 8x8 (or arbitrary) |
| Interface
|
FDL-PCI
|
| Mount |
** |
| Spectral
response |
350-1000
nm |
| Backfocus
|
17.5 mm |
| Cooling
|
Double
stage Peltier -50 °C DT |
| Shutter
|
Electromechanical
|
| Exposure
time |
From 0.01
to 9999 s |
| CCD temperature
control |
±0.1
°C |
| Settable
gains |
2 |
| BIAS Control
|
9 bit |
| Optical
window * |
fused silica
|
| Head weight
|
Starting
from 2000 g |
| Dimensions
|
Starting
from 130x125x150 mm |
| Auxiliary
port |
yes |
*
With double coating. On request, it can mount coating and/or double
coating windows with non parallel faces and windows in Berillium
with X-ray scintillator screen
** According to sensors
The standard
system includes:
- Assembled
unit in light alloy with threaded 42x0.75 input;
- 32 bit FDL-PCI interface;
- PC parallel link cable 2.5 m;
- software for image grab and processing under Windows 95/98/Me/XP/2000
with PCI interface (included);
- 230V power supply;
- case, manual with test report and 24 months of warranty.
Options
| FOL |
Serial
fiber optic link (fiber : 30 m, ST 62/125 mount).
Different lengths on demand |
| NIK-H |
Adapter
for Nikon lens |
| MIN-H |
Adapter
for 42x1 mm lens |
| M10 |
10 m parallel
link cable for FDL-PCI |
| TSC |
2 stage
Peltier + liquid heat exchange with external pump |
| LEX |
Liquid
heat exchanger |
| TTC |
Adapter
from 42x0.75 to C mount |
| ATE |
1/4" mounting
post |
All the data
reported are subject to change without prior warning.
DTA Scientific Instruments s.r.l. declines any responsibility for
errors that may occur in the specifications listed above.
Any
product of ours is CE compliant
|