CHROMA series is characterized by high sensitivity, fast image
downloading, reduced weight and dimensions. This includes the
cameras: CHROMA CX 3, CHROMA C3 Plus and CHROMA C4 (just as
PRELIMINARY).
All this kinds use Kodak KAF CCD sensors, displaying a wide
range of resolutions and pixels dimensions. The sensors are
front-illuminated, supported by an excellent Quantum Efficiency.
All of them (apart to those with the antiblooming option) can
offer a 100% fill factor, allowing one to shoot marvellous images.
Among the main application fields: Biology, Astronomy, Plasma
Physics, Semiconductor Physics.
The CHROMA C3 Plus is an upgrade for the CHROMA CX 3: it has
a better resolution due to 16 bit A/D Converter, a readout speed
>= 1Mpixel/s and, more, it is equipped of three integrated
peripherals:
1) two 32-bit generators able to yield impulses starting from
100 ns with internal time base, or from 20 ns with external
time base;
2) a 32-bit time delay with a 100 ns step with an internal time
base or 20 ns with external time base;
By means of these three devices, you can run an MCP, a LASER
or other external devices, as well as synchronize them to the
image acquisition.
Specifications
| MODEL
|
CCD
|
ARRAY
FORMAT (HxV) |
PIXEL
(HxV) um |
SIZE
(HxV) mm |
FWC
ke- |
DARK
CURRENT
e-/pix.s@-5 °C |
Qe
@ 450,550,650 nm |
FILL
FACTOR % |
| C3P-260E |
KAF-0261E |
512x512
|
20x20 |
10.2x10.2
|
500 |
12.5 |
35, 55,
58 |
100 |
| C3P-400LE
|
KAF-0401LE |
768x512
|
9x9 |
6.9x4.6
|
100 |
0.5 |
20, 35,
42 |
70 |
C3P-400E
C3P-400ME |
KAF-0402E
KAF-0402ME |
768x512
|
9x9 |
6.9x4.6
|
100 |
0.5 |
35, 55,
68
55, 67, 82 |
100 |
C3P-1600E
C3P-1600LE |
KAF-1602E
KAF-1602LE |
1536x1024 |
9x9 |
13.8x9.2
|
100
50 |
0.6
0.5 |
40, 50,
60
20, 35, 42 |
100
70 |
C3P-3200E
C3P-3200ME |
KAF-3200E
KAF-3200ME |
2184x1472
|
6.8x6.8
|
14.85x10.26
|
55 |
0.5 |
40, 52,
65
60, 75, 82 |
100 |
| C3P-5000C
|
KAF-5101CE |
2614x1966
|
6.8x6.8
|
17.8x13.4
|
40 |
0.5 |
31, 34,
31 |
68 |
C3P-6000E
C3P-6000LE |
KAF-6303E
KAF-6302LE |
3072x2048
3072x2034 |
9x9 |
27.65x18.48 |
100
50 |
0.5 |
40, 55,
64
9, 15, 13 |
100
70 |
"C"
stands for coloured sensor, "E" is the Enhanced (ITO) version
The values concerning Quantum Efficiency, Full-Well Capacity and
Dark Current have to be considered as typical.
| Readout
speed |
>= 1
Mpixel/s |
| Filter
wheel (opt.) |
8 or 6
positions according to different models |
| A/D Converter
|
16 bit
|
| Max total
noise (Typ.) |
10e- |
| Partial
CCD Reading |
Programmable
|
| Binning
|
From 1x1
to 8x8 or arbitrary |
| Interface
|
FDL-PCI
|
| Mount |
42x0.75
or C |
| Spectral
response |
350-1000
nm |
| Backfocus
|
26.8/17.5
mm |
| Cooling
|
Single
stage Peltier -40°C DT |
| Shutter
|
Electromechanical
|
| Exposure
time |
From 0.01
to 9999 s |
| CCD temperature
control |
±0.1°C |
| Settable
gains |
64 |
| BIAS Control
|
9 bit |
| Optical
window * |
BK7 with
2 AR |
| Weight
|
1600 g |
| Dimensions
|
155x134x62
mm |
| Auxiliary
port |
yes |
* With
double coating. On request, it can mount coating and/or double
coating windows with non parallel faces and windows in Berillium
with X-ray scintillator screen.
The standard
system includes:
-
assembled unit in light alloy with threaded 42x0.75 input;
- 32 bit FDL-PCI interface;
- 2.5 m PC parallel link cable;
- software for image grab and processing under Windows 95/98/Me/2000/XP
with PCI interface (included);
- 230V power supply;
- case, manual with test report and 24 months of warranty.
Options
| FOL |
Serial
fiber optic link (fiber : 30 m, ST 62/125 mount).
Different lengths on demand |
| RGB |
RGB interference
+ Blank, dia. 25 mm |
| WEL |
8-position
integrated filter wheel (6-position for CX6000), filter dia.
25mm |
| NIK-C |
Adapter
for Nikon lens |
| MIN-C |
Adapter
for 42x1 mm lens |
| PAR |
2.5 m standard
parallel port link cable |
| M10 |
10 m parallel
link cable for FDL-PCI |
| XTE |
1/4" mounting
post |
| TTC |
Adapter
from 42x0.75 to C mount |
| CPA |
Front panel
with C mount and backfocus (you can not fit the filter wheel)
|
All the data
reported are subject to change without prior warning.
DTA Scientific Instruments s.r.l. declines any responsibility for
errors that may occur in the specifications listed above.
Any
product of ours is CE compliant
data
sheet (.pdf)
|